Fully Stainless Steel Structure
The structure is described as fully stainless steel rather than the chamber alone, which addresses corrosion around the vessel as well as within it.
Separately Timed Drying
Time of sterilisation is 0 to 60 minutes and time for drying is separately 0 to 60 minutes, listed as two distinct rows rather than as one shared figure.
Dressing Drying Function
A drying function is fitted and described as suitable for medical dressing drying, which is what allows textile and wrapped loads to leave the chamber dry.
Three Chamber Capacities
Volumes of 150L, 200L and 280L are offered, at 440 by 1000mm, 515 by 1000mm and 600 by 1000mm respectively across the three models in this range.
Door Pressure Interlock
The door cannot be opened until chamber pressure has reduced to 0.027 MPa, and the unit cannot be started if the door does not close well before a cycle.
Safety Valve With Steam Recovery
The safety valve opens automatically when inner pressure exceeds 0.24 MPa, with the steam exhausted to the water tank rather than released into the department.
Low Water Protection
Power is cut off automatically and an alarm sounds when a water shortage occurs, so a dry run is prevented and not merely signalled to whoever is nearby.
Shared Operating Parameters
Working pressure is 0.22 MPa and working temperature 134 degrees on every model, with adjustment from 40 to 134 degrees and heat average within plus or minus 2 degrees.