The LM-7 serves glazing workshops, optical laboratories, dispensing counters and refraction rooms. Glazing technicians, dispensing opticians, and optometrists use it to centre and mark lenses for glazing, neutralise a patient's current spectacles, confirm finished pairs before they leave the workshop, and measure contact lenses on the supplied stand.
Prism work is the LM-7's natural territory. With the prescribed prism keyed in, the layout screen shows the lens position that delivers it, in Cartesian or polar form, so the technician centres and marks the lens to that position before edging. Workshops that handle prism regularly gain a guided bench step. Prism up to 20 Δ falls within the measuring range.
Practices that keep their records digital are the other natural fit. Results leave through RS-232C, USB, a card reader or LAN, and wireless LAN can be added, so readings move straight into practice systems. Measurement takes about 0.1 seconds across 108 points in the nosepiece, under 538 nm green light close to the ISO standard wavelength, so high-index lenses need no Abbe value.
At the dispensing counter, PD is read against the on-screen scale while the patient's frame is held in front of it, and UV-A transmission at 365 nm can be compared between two lens types. Progressive lenses are recognised as they land on the nosepiece. Upkeep centres on red ink cartridges for marking, with blue cartridges as an option, and a dust cover protects the instrument between sessions. The unit runs on 100–240 V and draws up to 50 VA.